Traduciendo...
STAS1 Off
Safety Node A Power Supply Not Present
Kuka · Sunrise Cabinet Med Controller
¿Qué significa STAS1 Off ?
The STAS1 (Safety node A) Orange LED is OFF, indicating no supply voltage is present for this safety node. This absence of power prevents Safety Node A from operating, which could compromise the overall safety functionality of the system.
Causas comunes
- The power supply for Safety Node A (STAS1) is off.
- The STAS1 safety device has lost its power connection.
- No voltage detected on the STAS1 safety module's power input.
Pasos de reparación & Lista de verificación
Haga clic en los pasos para seguir su progreso.
- 1
Check fuse F17.3 on the CCU_SR for continuity and proper rating.
- 2
If the PWR/3.3V LED (Item 6, Page 98) lights up, indicating main 3.3V power is present, exchange the CCU_SR module, as the safety node's power supply is likely faulty internally.
Datos técnicos verificados. Última actualización: March 2026
Fallas relacionadas