PMB_SR Microcontroller Internal Fault
Kuka · Sunrise Cabinet Med Controller
What does STA1 (PMB_SR) Flashing mean?
The STA1 (PMB_SR) μC USB LED is flashing at a rate other than 1 Hz (normal) or 10 Hz (boot phase), indicating an internal fault within the PMB_SR microcontroller. This suggests the microcontroller is not operating correctly, possibly due to a core component malfunction or a corrupt firmware state. The fault might be related to the power supply to the module or internal processing errors.
Common Causes
- The PMB_SR microcontroller has encountered an internal processing error, indicated by flashing status.
- Internal fault within the power module board microcontroller.
- The PMB_SR module's internal software or hardware has failed.
Repair Steps & Checklist
Click steps to track your progress.
- 1
Check the infeed power supply at connector X1.
- 2
Verify the status of the PWR/5V LED. If the PWR/5V LED lights up, exchange the CCU_SR module, as the internal fault persists despite power being present.
- 3
If the PWR/5V LED is not lit, troubleshoot the 5V power supply before considering CCU_SR module replacement.