PWR/3.3V Off (CIB_SR)

CIB_SR 3.3V Power Supply Not Present

Kuka · Sunrise Cabinet Med Controller

What does PWR/3.3V Off (CIB_SR) mean?

The PWR/3.3V Green LED for the CIB_SR module is OFF, indicating a lack of 3.3V supply voltage to the module. This absence of critical operating voltage prevents the CIB_SR module from functioning correctly and could result in the non-operation of interconnected I/O nodes.

Common Causes

  • The 3.3V power supply to the CIB_SR module is not present.
  • Absence of 3.3V power on the CIB_SR communication interface board.
  • The CIB_SR module is not receiving its required 3.3V supply voltage.

Repair Steps & Checklist

Click steps to track your progress.

  1. 1

    Check fuse F17.3 on the CCU_SR for continuity and proper rating.

  2. 2

    Verify the presence and correct seating of jumper plug X308.

  3. 3

    Check fuse F308 on the CCU_SR for continuity and proper rating.

  4. 4

    If an external supply is used via X308, check the external power supply to ensure it provides the rated voltage of 24 V.