PWR/1.2V Off (CIB_SR)

CIB_SR 1.2V Power Supply Not Present

Kuka · Sunrise Cabinet Med Controller

What does PWR/1.2V Off (CIB_SR) mean?

The PWR/1.2V Green LED for the CIB_SR module is OFF, indicating a lack of 1.2V supply voltage to the module. This absence of critical operating voltage prevents the CIB_SR module from functioning correctly and could result in the non-operation of interconnected I/O nodes.

Common Causes

  • The 1.2V power supply to the CIB_SR module is not present.
  • The CIB_SR module requires 1.2V power, which is currently absent.
  • Failure or absence of the 1.2V power rail for the CIB_SR communication board.

Repair Steps & Checklist

Click steps to track your progress.

  1. 1

    Check fuse F17.3 on the CCU_SR for continuity and proper rating.

  2. 2

    Verify the presence and correct seating of jumper plug X308.

  3. 3

    Check fuse F308 on the CCU_SR for continuity and proper rating.

  4. 4

    If an external supply is used via X308, check the external power supply to ensure it provides the rated voltage of 24 V.

Browse all codes in this manual (21)
Verified technical data. Last updated: March 2026

Related Faults

Source: Kuka Sunrise Cabinet Med Controller