Übersetzen...
STAS1 Flashing

Safety Node A Internal Fault

Kuka · Sunrise Cabinet Med Controller

Was bedeutet STAS1 Flashing ?

The STAS1 (Safety node A) Orange LED is flashing at a rate other than 1 Hz (normal) or 10 Hz (boot phase), indicating an internal fault code within Safety Node A. This suggests a problem with the internal logic, communication, or processing of this safety node, potentially affecting safety chain integrity and preventing safe operation.

Häufige Ursachen

  • Internal fault within the Safety Node A (STA1).
  • Possible hardware failure on the safety controller.
  • Corrupted firmware in the safety module.
  • Overload or short circuit detected on a safety input/output.

Reparaturschritte & Checkliste

Klicken Sie auf Schritte, um Ihren Fortschritt zu verfolgen.

  1. 1

    Check cabling at connectors X309, X310, and X312 for proper connection, integrity, and any signs of damage.

  2. 2

    For test purposes, disconnect the cables at X309, X310, and X312, then switch the controller off and back on again to reset the module.

  3. 3

    If the error recurs after checking cabling and power cycling, exchange the CCU_SR module.

Alle Codes in diesem Handbuch durchsuchen (21)
Verifizierte technische Daten. Zuletzt aktualisiert: March 2026

Verwandte Fehlercodes

Quelle: Kuka Sunrise Cabinet Med Controller