PWR/2.5V Off (CIB_SR)

CIB_SR 2.5V Power Supply Not Present

Kuka · Sunrise Cabinet Med Controller

Was bedeutet PWR/2.5V Off (CIB_SR) ?

The PWR/2.5V Green LED for the CIB_SR module is OFF, indicating a lack of 2.5V supply voltage to the module. This absence of critical operating voltage prevents the CIB_SR module from functioning correctly and could result in the non-operation of interconnected I/O nodes.

Häufige Ursachen

  • The 2.5V power supply to the CIB_SR module is not present.
  • No 2.5V power detected on the CIB_SR board.
  • The CIB_SR communication interface board is missing its 2.5V power input.

Reparaturschritte & Checkliste

Klicken Sie auf Schritte, um Ihren Fortschritt zu verfolgen.

  1. 1

    Check fuse F17.3 on the CCU_SR for continuity and proper rating.

  2. 2

    Verify the presence and correct seating of jumper plug X308.

  3. 3

    Check fuse F308 on the CCU_SR for continuity and proper rating.

  4. 4

    If an external supply is used via X308, check the external power supply to ensure it provides the rated voltage of 24 V.