PWR_+12V5 Off
Brake Chopper Power Supply Not Present
Kuka · Sunrise Cabinet Med Controller
Was bedeutet PWR_+12V5 Off ?
The PWR_+12V5 Green LED is OFF, indicating that the brake chopper power supply is not present. This condition means the 12.5V power rail required for the brake chopper operation is missing, which could impact the robot's ability to dissipate regenerative energy during braking. Without this power, the brake chopper cannot function.
Häufige Ursachen
- The +12.5V power supply to the brake chopper circuit is not present.
- No voltage detected on the +12.5V line supplying the brake chopper.
- Failure in the power source for the brake chopper control (e.g., fuse, regulator).
Reparaturschritte & Checkliste
Klicken Sie auf Schritte, um Ihren Fortschritt zu verfolgen.
- 1
Check the integrity and ensure proper seating of connector X500.
- 2
Check the connector on the low-voltage power supply unit for proper connection and any visible damage.